Setting the gfc7001t concentration alarm limits, Teledyne analytical instruments 137, Alarm status limit set point – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 137

Advertising
background image

SetupTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

137

a Single Pole Double Throw relay output accessible via the alarm output connector on
the instrument’s back panel (See Section 3.3.1.4). If the CO concentration measured by
the instrument rises above that limit, the alarm‘s status output relay is closed.
The default settings for ALM1 and ALM2 are:

Table 5-10:

CO Concentration Alarm Default Settings

ALARM STATUS

LIMIT

SET

POINT

1

alm1 Disabled

100

ppm

alm2 Disabled

300

ppm

1

Set points listed are for PPM. Should the reporting range units of measure be changed (See

Section 5.4.3) the analyzer will automatically scale the set points to match the new range unit
setting.

Note

To prevent the concentration alarms from activating during span
calibration operations ensure that the CAL or CALS button is pressed
prior to introducing span gas into the analyzer.

5.10.1. SETTING THE GFC7001T CONCENTRATION ALARM LIMITS

To enable either of the CO concentration alarms and set the limit points, press:

SETUP X.X

CO ALRM 2: OFF

ON

ENTR EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS DIAG

ALRM

EXIT

DIAG FCAL CO ALARM 2=300.00 PPM

3

0

0

.0

0

ENTR

EXIT

Toggle these

buttons to Set

the alarm point

EXIT discards the new

setting

ENTR accepts the

new setting

SAMPLE

RANGE=50.0 PPM CO= XX.XX

<TST

TST> CAL

SETUP

SETUP X.X

CO ALRM 2, DISALBED

NEXT

EDIT

PRNT EXIT

Toggle this button to

enable/disable

the alarm

Continue pressing NEXT until the desired

Alarm is selected