Table 11-1: GFC7001T/GFC7001TM Maintenance Schedule

ITEM ACTION FREQ

CAL

CHECK

REQ’D

MANUAL

DATE PERFORMED

Particulate

Filter

Replace

Weekly or As

Needed

No

Verify Test

Functions

Record and

Analyze

Weekly or after

any

Maintenance or

Repair

No

Perform Flow

Check

Check Flow

Annually

No

Perform

Leak Check

Verify Leak

Tight

Annually or

after any

Maintenance or

Repair

No

Pneumatic

lines

Examine and

Clean

As Needed

Yes if

cleaned

Cleaning Clean As

Needed

Only if

cover

removed