Calibration Procedures Teledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
213
EXIT returns
to the SETUP
Menu.
SETUP X.X
DURATION:15.0 MINUTES
<SET SET>
EDIT
EXIT
SETUP X.X
DURATION 15.0MINUTES
3 0 .0
ENTR
EXIT
SETUP X.X
DURATION:30.0 MINUTES
<SET
SET>
EDIT
EXIT
Toggle buttons to set
duration for each iteration
of the sequence:
Set in Decimal minutes
from 0.1 – 60.0.
SETUP X.X
CALIBRATE: OFF
<SET SET>
EDIT
EXIT
SETUP X.X CALIBRATE: OFF
ON
ENTR
EXIT
SETUP X.X
CALIBRATE: ON
<SET SET> EDIT
EXIT
Toggle button between
Off and ON.
SETUP X.X
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
Display show:
SEQ 2) ZERO–SPAN, 2:00:30
Sequence
Delta Time
MODE Delta Days
CONTINUED FROM PREVIOUS PAGE
DELTA TIME
9.4.1.1. AUTOCAL WITH AUTO OR DUAL REPORTING RANGES MODES
SELECTED
If the GFC7001T/GFC7001TM Analyzer is set for either the Dual or Auto reporting
range modes, the following three steps will appear at the beginning of the AutoCal setup
routine:
SETUP X.X
RANGE TO CAL: HIGH
<SET
EDIT
EXIT
EXIT returns to the
PRIMARY SETUP
Menu.
SETUP X.X
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
SETUP X.X
RANGE TO CAL: LOW
LOW
HIGH
ENTR
SETUP
SETUP X.X
RANGE TO CAL: LOW
<SET
EDIT
EXIT
Note
In order to automatically calibrate both the HIGH and LOW ranges, you
must set up a separate sequence for each.