Sensor calibration procedure – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 222

Advertising
background image

Calibration Procedures Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

222

9.7.2. CO

2

SENSOR CALIBRATION PROCEDURE

Presented here are first the setup and then the calibration steps for the CO

2

Sensor.

9.7.2.1. CO2 PNEUMATICS CONNECTIONS

The pneumatic connections for calibrating are as follows

Figure 9-8:

CO

2

Sensor Calibration Set Up

CO

2

SENSOR ZERO GAS: TAI recommends using pure N

2

when calibration the zero

point of your CO

2

sensor option.

CO

2

SENSOR SPAN GAS: TAI recommends using 16% CO

2

in N

2

when calibration

the span point of your CO

2

sensor option (Table 3-12) is 20%.

9.7.2.2. SET CO2 SPAN GAS CONCENTRATION:

Set the expected CO

2

span gas concentration.

This should be equal to the percent concentration of the CO

2

span gas of the selected

reporting range (default factory setting = 12%).