BENCH TEMP WARNING

The temperature of the optical bench is outside the specified limits.

BOX TEMP WARNING

The temperature inside the chassis is outside the specified limits.

CANNOT DYN SPAN

2

Remote span calibration failed while the dynamic span feature was set to turned on.

CANNOT DYN ZERO

3

Remote zero calibration failed while the dynamic zero feature was set to turned on.

CONC ALRM1 WARNING

1

Concentration alarm 1 is enabled and the measured CO level is ≥ the set point.

CONC ALRM2 WARNING

1

Concentration alarm 2 is enabled and the measured CO level is ≥ the set point.

CONFIG INITIALIZED

Configuration storage was reset to factory configuration or erased.

DATA INITIALIZED

DAS data storage was erased.

O

2

CELL TEMP WARN

2

O

2

sensor cell temperature outside of warning limits.

PHOTO TEMP WARNING

The temperature of the IR photo detector is outside the specified limits.

REAR BOARD NOT DET

The CPU is unable to communicate with the motherboard.

RELAY BOARD WARN

The firmware is unable to communicate with the relay board.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SAMPLE PRESS WARN

Sample gas pressure outside of

operational parameters.

SAMPLE TEMP WARN

The temperature of the sample gas is outside the specified limits.

SOURCE WARNING

The IR source may be faulty.

SYSTEM RESET

1

The computer was rebooted.

WHEEL TEMP WARNING

The Gas Filter Correlation Wheel temperature is outside the specified limits.

1

Alarm warnings only present when optional alarm package is activated.

2

Only enabled when the optional O

2

Sensor is installed.

To view and clear warning messages: