Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 70

Advertising
background image

Getting StartedTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

70

 At least 0.2m long;
 No more than 2m long and;
 Vented outside the shelter or immediate area surrounding the instrument.
A similar vent line should be connected to the VENT SPAN outlet on the back of the
analyzer.

E

XHAUST

O

UTLET

Attach an exhaust line to the analyzer’s EXHAUST outlet fitting. The exhaust line
should be:
 PTEF tubing; minimum O.D ¼”;
 A maximum of 10 meters long;
 Vented outside the analyzer’s enclosure.

3.3.2.9. PNEUMATIC LAYOUT FOR ZERO SCRUBBER/ AMBIENT SPAN OPTION

Figure 3-27:

Internal Pneumatic Flow OPT 50H – Zero Scrubber/Ambient Span

Table 3-11: Zero/Span Valve Operating States for Option 50H

MODE VALVE

CONDITION

SAMPLE

(Normal

State)

Sample/Cal

Open to SAMPLE inlet

Zero/Span

Open to ZERO AIR scrubber

ZERO CAL

Sample/Cal

Open to ZERO/SPAN valve

Zero/Span

Open to ZERO AIR scrubber

SPAN CAL

Sample/Cal

Open to ZERO/SPAN valve

Zero/Span

Open to PRESSURE SPAN inlet

Advertising