Theory of Operation Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

293

IR unaffected by N

2

in Measurement Cell of

the GDC Wheel and no additional CO in the

Sample Chamber

IR affected by CO in Reference Cell

with no interfering gas in the

Sample Chamber

IR shinning through Measurement Cell of

the GDC Wheel is reduced by additional CO

in the Sample Chamber

IR shining through Reference Cell is

also reduced by additional CO in the

Sample Chamber, but to a lesser extent

M/R

is reduced

CO REF

CO MEAS

Figure 13-4:

Affect of CO in the Sample on CO MEAS & CO REF

Once the GFC7001T/GFC7001TM has computed this ratio, a look-up table is used, with
interpolation, to linearize the response of the instrument. This linearized concentration
value is combined with calibration SLOPE and OFFSET values to produce the CO
concentration which is then normalized for changes in sample pressure.

I

NTERFERENCE AND

S

IGNAL TO

N

OISE

R

EJECTION

If an interfering gas, such as H

2

O vapor is introduced into the sample chamber, the

spectrum of the IR beam is changed in a way that is identical for both the reference and
the measurement cells, but without changing the ratio between the peak heights of CO
MEAS
and CO REF. In effect, the difference between the peak heights remains the
same.