6.7.3.4. SETTING HESSEN PROTOCOL STATUS FLAGS

Teledyne’s implementation of Hessen protocols includes a set of status bits that the
instrument includes in responses to inform the host computer of its condition. Each bit
can be assigned to one operational and warning message flag. The default settings for
these bit/flags are:

Table 6-6: Default Hessen Status Flag Assignments

STATUS FLAG NAME

DEFAULT BIT ASSIGNMENT

WARNING FLAGS

SAMPLE FLOW WARNING

0001

BENCH TEMP WARNING

0002

SOURCE WARNING

0004

BOX TEMP WARNING

0008

WHEEL TEMP WARNING

0010

SAMPLE TEMP WARN

0020

SAMPLE PRESS WARN

0040

INVALID CONC

(The Instrument’s Front Panel Display Will Show The
Concentration As “Warnings”)

0080

OPERATIONAL FLAGS

1

Instrument OFF 0100

In MANUAL Calibration Mode

0200

In ZERO Calibration Mode

4

0400

In

O

2

Calibration Mode (if

O

2

sensor installed )

2,4

0400

In

CO

2

Calibration Mode (if

CO

2

sensor installed )

2,4

0400

In SPAN Calibration Mode

0800

UNITS OF MEASURE FLAGS

UGM

0000

MGM

2000

PPB

4000

PPM

6000

SPARE/UNUSED BITS

1000, 8000

UNASSIGNED FLAGS (0000)

AZERO WARN

2

DCPS WARNING

CANNOT DYN SPAN

2

REAR BOARD NOT DET

CANNOT DYN ZERO

3

SYNC

WARNING

1

CONC ALARM 1

3

SYSTEM RESET

1

CONC ALARM 2

3

1

These status flags are standard for all instruments and should probably not be modified.

2

Only applicable if the optional internal span gas generator is installed.

3

Only applicable if the analyzer is equipped with an alarm options.

3

It is possible to assign more than one flag to the same Hessen status bit. This allows the grouping of
similar flags, such as all temperature warnings, under the same status bit.

Be careful not to assign conflicting flags to the same bit as each status bit will be triggered if any of the
assigned flags is active.