Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 117

Advertising
background image

SetupTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

117

Table 5-5:

DIAG - Analog I/O Functions

SUB MENU

OUTPUT

CHANNEL

FUNCTION

AOUT

CALIBRATED

ALL

Initiates a calibration of the A1, A2, A3 and A4 analog output channels that
determines the slope and offset inherent in the circuitry of each output.
These values are stored and applied to the output signals by the CPU
automatically.

CONC_OUT_1

A1

Sets the basic electronic configuration of the A1 output (CO Concentration).
There are four options:

RANGE

1

: Selects the signal type (voltage or current loop) and level of the

output.

REC OFS

1

: Allows them input of a DC offset to let the user manually adjust

the output level.

AUTO CAL: Enables / Disables the AOUT CALIBRATED feature.
CALIBRATED: Performs the same calibration as AOUT CALIBRATED,

but on this one channel only.

CONC_OUT_2

A2

 Same as for CONC_OUT_1 but for analog channel A2.

CONC_OUT_3

A3

 Same as for CONC_OUT_1 but for analog channel A3 but only if either the

optional O

2

or CO

2

sensors are installed.

TEST OUTPUT

A4

 Same as for CONC_OUT_1 but for analog channel A4 (TEST CHANNEL).

AIN

CALIBRATED

N/A

Initiates a calibration of the A-to-D Converter circuit located on the
Motherboard.

XIN1

.
.
.

XIN8

For each of 8 external analog input channels, shows the gain, offset,
engineering units, and whether the channel is to show up as a Test function.

1

Any changes made to RANGE or REC_OFS require recalibration of this output.