Teledyne analytical instruments 234 – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 234

Advertising
background image

EPA Calibration Protocol Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

234


SPAN CAL M

A1:CONC1=50 PPM CO = XXXX

< TST TST >

ZERO

SPAN

CONC

EXIT

Wait until

STABIL falls

below 0.2 PPM

(for M300E).

This may take
several minutes.

SAMPLE

COSTB=XXXX PPB

CO=XXXX

< TST TST > CAL

CALZ

CALS

SETUP

Record the CO

reading as

displayed on the

instrument’s front

panel.

Press EXIT to

Return to the

Main SAMPLE

Display.

ACTION:

Allow Calibration Gas diluted to

proper concentration for

Midpoint N+1 to enter the sample

port.

ACTION:

Allow calibration gas diluted to proper concentration for

Midpoint N to enter the sample port

SAMPLE A1:CONC1=50 PPM CO = XXXX

< TST

TST

>

CAL

SETUP

SAMPLE

CO STB=XXXX PPB

CO=XXXX

< TST TST >

CAL

SETUP

Set the Display to show the

COSTB test function.

This function calculates the

stability of the CO

measurement.

Plot the analyzer responses versus the corresponding calculated concentrations to obtain
a calibration relationship. Determine the best-fit straight line (y = mx + b) determined
by the method of least squares.
After the best-fit line has been drawn, determine whether the analyzer response is linear.
To be considered linear, no calibration point should differ from the best-fit line by more
than

2% of full scale.

If carried out carefully, the checks described in this section will provide reasonable
confidence that the GFC7001T is operating properly. Checks should be carried out at
least every 3 months as the possibility of malfunction is always present.
If the linearity error is excessive and cannot be attributed to outside causes, check the
GFC7001T system for:
 Sample pressure higher than ambient – pressurized sample gas
 Leaks
 Correct flow
 Miscalibrated span gas tanks or bad zero gas
 Miscalibrated sample pressure transducer

Advertising