ProSoft Technology 5105-103M-PDPS User Manual

Page 78

Advertising
background image

Diagnostics and Troubleshooting

5105-103M-PDPS ♦ ProLinx Gateway

User Manual

IEC 60870-5-103 Master to PROFIBUS Slave Gateway

Page 78 of 144

ProSoft Technology, Inc.

November 24, 2010

2.9.8 Stopping the Data Analyzer

Press [S]

to stop the data analyzer. Use this option to freeze the display so the

data can be analyzed. To restart the analyzer, press [B].

Important: When in analyzer mode, program execution will slow down. Only use this tool during a
troubleshooting session. Before disconnecting from the Config/Debug port, please press [S]

to stop

the data analyzer, and then press [M]

to return to the main menu. This action will allow the module

to resume its normal high speed operating mode.

2.9.9 Data Analyzer Tips

From the main menu, press [A]

for the "Data Analyzer". You should see the

following text appear on the screen:

After the "Data Analyzer" mode has been selected, press [?]

to view the Data

Analyzer menu. You will see the following menu:

From this menu, you can select the "Port", the "format", and the "ticks" that you

can display the data in.
For most applications, HEX is the best format to view the data, and this does

include ASCII based messages (because some characters will not display on

HyperTerminal and by capturing the data in HEX, we can figure out what the

corresponding ASCII characters are supposed to be).
The Tick value is a timing mark. The module will print a _TT for every xx

milliseconds of no data on the line. Usually 10milliseconds is the best value to

start with.

Advertising
This manual is related to the following products: