KEYENCE CB-EP100 User Manual

Page 100

Advertising
background image

- EtherNet/IP Network Unit CB-EP100 User’s Manual (LJ-V7000 Edition) -

3-4 Message Communication

3-76

3

Communi

cat

in

g wit

h the

L

J

-V700

0 Ser

ies

• Responses (when failed)

For details on the attribute ID, settings, general status, and additional status of each
parameter, refer to

"CB Object (Class ID:6BH)" (Page 3-40) and

"Attributes"

(Page 3-44).

(3) Changing programs

Message communication changes the programs of a measuring instrument
connected to the CB-EP100.
Service code 4BH is used.
The following settings are required to use this object.

• Service code
• Class ID
• Instance ID
• Service Data

• Command

• Response (when successful)

• Response (when failed)

Item

Description

Data Example

General status

A general status for the service is returned

0EH

Additional status

An additional status is returned.

-

Service response data

There is no service response data.

-

Item

Description

Data Example

Service code

"Program number change (4BH) " is specified (fixed
value).

4BH

Class ID

CB Object (6BH) is specified (fixed value).

6BH

Instance ID

01H becomes a fixed value.

01H

Attribute ID

There is no attribute ID.

-

Service data

Specifies the number of the program to change to.

00000000H-

0000000FH

Item

Description

Data Example

General status

Message communication returns status (00H) to notify
normal termination.

00H

Service response data

There is no service data.

-

Item

Description

Data Example

General status

Message communication returns a general status.

14H

Additional status

Returns the additional status.

-

Service response data

There is no service data.

-

Advertising