3-4 Message Communication

3-75

3

Communi

cat

in

g wit

h the

L

J

-V700

0 Ser

ies

- EtherNet/IP Network Unit CB-EP100 User’s Manual (LJ-V7000 Edition) -

• Responses (when failed)

For details on the attribute ID, settings, general status, and additional status of each
parameter, refer to

"CB Object (Class ID:6BH)" (Page 3-40) and

"Attributes"

(Page 3-44).

(2) Writing setting values to a measuring instrument

Message communication writes parameters to the CB-EP100 or to measuring
instrument connected to the CB-EP100.
Service code 10H is used.
The following settings are required to use this object.

• Service Code
• Class ID
• Instance ID
• Attribute ID
• Service Data

With this service, setting values will not be reflected in actual operation
after they have been written unless a request to reflect the setting writing
area is used.

• Commands

• Responses (when successful)

Item

Description

Data Example

General status

A general status for the service is returned

14H

Additional status

An additional status is returned.

-

Service response data

There is no service response data.

-

Item

Description

Data Example

Service Code

Designates "Parameter write (10H)" (fixed value).

10H

Class ID

Designates CB Object(6BH) (fixed value).

6BH

Instance ID

Designates the ID number.

C8H

Attribute ID

Designates the attribute ID to be written from "Attribute".

991H(2449)

Service Data

Writes the setting.

00000000H

Item

Description

Data Example

General status

A status (00H) is returned that indicates normal
completion of message communication. (The written
parameters are reflected in the measuring instrument at
the status (00H) return timing.)

00H

Service response data

There is no service response data.

-

Point