KEYENCE CB-EP100 User Manual
Page 105
3-4 Message Communication
3-81
3
Communi
cat
in
g wit
h the
L
J
-V700
0 Ser
ies
- EtherNet/IP Network Unit CB-EP100 User’s Manual (LJ-V7000 Edition) -
(7) Updating setting write area
Message communication updates the contents of the setting write area of the
measuring instrument connected to the CB-EP100 with the contents of the specified
hierarchy level. Service code 4FH is used.
The following settings are required to use this object.
• Service code
• Class ID
• Instance ID
• Service Data
• Commands
• Response (when successful)
• Response (when failed)
Item
Description
Data Example
Service code
Specify "update setting write area (4FH)" (fixed value).
4FH
Class ID
CB Object (6BH) is specified (fixed value).
6BH
Instance ID
01H becomes a fixed value.
01H
Attribute ID
There is no data.
-
Service data
Update source area (0 = setting area currently in
operation, 1 = save area)
00000000H or
00000001H
Item
Description
Data Example
General status
Message communication returns status (00H) to notify
normal termination.
00H
Service response data
There is no service response data.
-
Item
Description
Data Example
General status
Message communication returns a general status.
05H
Additional status
Returns the additional status.
-
Service response data
There is no service response data.
-