Attribute ids supported, 5) control input to a measuring instrument – KEYENCE CB-EP100 User Manual

Page 102

Advertising
background image

- EtherNet/IP Network Unit CB-EP100 User’s Manual (LJ-V7000 Edition) -

3-4 Message Communication

3-78

3

Communi

cat

in

g wit

h the

L

J

-V700

0 Ser

ies

(5) Control input to a measuring instrument

Message communication implements control input such as auto zero and reset for
measuring instrument connected to the CB-EP100. Service code 4DH is used.
The following settings are required to use this object.

• Service Code

• Class ID

• Instance ID

• Attribute ID

• Service Data

Before executing control input, make sure that the measuring instrument
can execute control input. For the confirmation method, refer to LJ-V7000
Series User's Manual . After executing control input, confirm the successful
completion of the control input.

• Attribute IDs supported by the instance service 4DH (control input)

Processing performed by the instance service 4DH (control input) depends on the
selected attribute ID.
The table below describes the IDs that can be selected. (The instance ID is fixed
to 01H)

Attribute ID

Name

Description

Service

data

006EH

(110)

Trigger

Indicates measurement trigger input.

None

006FH

(111)

Auto-zero

Turns auto-zero ON/OFF. The auto-zero
reference value is the measurement value
when the auto-zero function is turned on. The
measurement values are updated every time
the auto-zero function is turned on. The
reference value is reset when the auto-zero
function is turned off. This function is enabled
only in high-performance mode.

Target OUT
information

*1

0070H

(112)

Timing

Turns timing input on or off. This function is
enabled only in high-performance mode.

Target OUT
information

*1

0071H

(113)

Reset

Performs reset input. Reset input ON

⇒ OFF is

ignored. This function is enabled only in high-
performance mode.

Target OUT
information

*2

0072H

(114)

Memory clear

Clears the internal memory. The internal
memory is an area used for buffering profiles
in high-speed mode and used as a storage
buffer in high-performance operating mode.

None

0073H

(115)

Batch start

Indicates that batch measurements have
started.

None

0074H

(116)

Batch end

Indicates that batch measurements have
ended.

None

Point