3-4 Message Communication

3-31

3

Communi

cat

in

g wit

h the

L

J

-V700

0 Ser

ies

- EtherNet/IP Network Unit CB-EP100 User’s Manual (LJ-V7000 Edition) -

(5) Inputting control data

Message communication implements control input for a measuring instrument
connected to the CB-EP100.
In the following example, the auto-zero operation is performed.

"(5) Control input to a measuring instrument" (Page 3-78)

(6) Requesting to reflect setting write area

Message communication reflects the setting values of the setting write area of the
measuring instrument connected to the CB-EP100 up to the settings of the
specified hierarchy level.

"(6) Requesting to reflect setting write area" (Page 3-80)

(7) Updating the setting write area

Message communication updates the contents of the setting write area of the
measuring instrument connected to the CB-EP100 with the contents of the
specified hierarchy level.

"(7) Updating setting write area" (Page 3-81)

Command

Response

CB-EP100

Scanner

<Details of command>
Service code: 4DH
Class ID: 6BH
Instance ID: 01H
Attribute ID: 6FH
Service data: Target OUT
information

Measuring
instrument

Command

Response

CB-EP100

Scanner

<Details of command>
Service code: 4EH
Class ID: 6BH
Instance ID: 01H
Attribute ID: None
Service data: 0000 0001H

Measuring
instrument

The setting values of the setting
write area are reflected up to
the settings of save area.

Command

Response

CB-EP100

Scanner

<Details of command>
Service code: 4FH
Class ID: 6BH
Instance ID: 01H
Attribute ID: None
Service data: 0000 0000H

Measuring
instrument

The setting values of the setting write
area are replaced with the settings of
the active measurement area.