- EtherNet/IP Network Unit CB-EP100 User’s Manual (LJ-V7000 Edition) -

3-4 Message Communication

3-84

3

Communi

cat

in

g wit

h the

L

J

-V700

0 Ser

ies

(10)Setting the date and time

Message communication sets the date and time of a measuring instrument
connected to the CB-EP100.
Service code 58H is used.
The following settings are required to use this object.

• Service code

• Class ID

• Instance ID

• Service Data

• Commands

• Response (when successful)

• Response (when failed)

Item

Description

Data Example

Service code

Specify "set date and time (58H)" (fixed value).

58H

Class ID

CB Object (6BH) is specified (fixed value).

6BH

Instance ID

01H becomes a fixed value.

01H

Attribute ID

There is no attribute ID.

-

Service data

Specifies the date and time settings.

Data length, year,

month, day, hour,

minute, second

Item

Description

Data Example

General status

Message communication returns status (00H) to notify
normal termination.

00H

Service response data

There is no service response data.

-

Item

Description

Data Example

General status

Message communication returns a general status.

05H

Additional status

Returns the additional status.

-

Service response data

There is no service response data.

-