3-4 Message Communication

3-45

3

Communi

cat

in

g wit

h the

L

J

-V700

0 Ser

ies

- EtherNet/IP Network Unit CB-EP100 User’s Manual (LJ-V7000 Edition) -

Instance ID = 1
Data unique to a measuring instrument (LJ-V) connected to the CB-EP100 can be
obtained. The setting values obtained are the settings of the setting area currently in
operation for the program number at that point in time.

Service

code

(hexadecimal)

Attribute

ID

(hexadecimal)

Name

Description and service data

Attribute

Data type

0EH

65H

Setting values are
being collected

Indicates whether or not the latest set-
ting information of measuring instru-
ment is currently being collected. 0
indicates that the information is not
being collected.

R

DWORD

66H

Save memory
status

Indicates whether data is being writ-
ten to the save memory. 0 indicates
that data is not being written.

DWORD

67H

Error code

Stores the error codes of connected a
measuring instrument.

UINT

68H

Warning code

Stores the warning codes of con-
nected a measuring instrument.

UINT

69H

-

6DH

Reserved

4DH

6EH

Trigger

For information on detailed formats,
refer to

"Attribute IDs supported

by the instance service 4DH (control
input)" (Page 3-78).

-

Page 3-78

6FH

Auto-zero

70H

Timing

71H

Reset

72H

Memory clear

73H

Batch start

74H

Batch end

75H

Storage start

76H

Storage end

78H

-

2FFH

Reserved

0EH

300H

System Status

For information on detailed formats,

refer to

"Monitor Data (156 Words

(312 Bytes)) Assembly Instance
(Instance ID): 64H" (Page 3-9)

R

WORD

301H

Measuring in
strument status

302H

Valid status

303H

Judgment
standby status

304H

Alarm status

305H

-

307H