Chapter 4 Remote Interface Reference

Calibration Commands

98

CALibration:VOLTage:LEVel {MINimum | MIDdle|MAXimum}

This command can only be used after calibration is unsecured and the output

state is ON. It sets the power supply to a calibration point that is entered with

CAL:VOLT

command. During calibration, three points must be entered and the

low-end point (MIN) must be selected and entered first.

CALibration:VOLTage:PROTection

This command calibrates the overvoltage protection circuit of the power

supply. It takes about 7 seconds to execute the command. The calibration must

be unsecured and the output be opened before calibrating the overvoltage

protection circuit. The power supply automatically performs the calibration

and stores the new overvoltage constant in

nonvolatile

memory.

Notice that

voltage calibration precedes before sending this command.