Chapter 5 Error Messages

Calibration Errors

129

5

Calibration Errors

The following errors indicate failures that may occur during a calibration.

Refer to the

Service Guide

for more information.

701

Cal security disabled by jumper

The calibration security feature has been disabled with a jumper inside the

power supply. When applicable, this error will occur at power-on to warn you

that the power supply is unsecured.

702

Cal secured

The power supply is secured against calibration.

703

Invalid secure code

An invalid calibration security code was received when attempting to unsecure

or secure the power supply. You must use the same security code to unsecure

the power supply as was used to secure it, and vice versa. The security code

may contain up to 12 alphanumeric characters. The first character must be a

letter.

704

Secure code too long

A security code was received which contained more than 12 characters.

705

Cal aborted

A calibration in progress is aborted when you press any front-panel key, send

a device clear, or change the local/remote state of the instrument.

708

Cal output disabled

Calibration is aborted by sending

OUTP OFF

command during calibrating a

output.

712

Bad DAC cal data

The specified DAC calibration values (

CAL:VOLT

or

CAL:CURR

) are out of

range. Note that the new calibration constants are not stored in the non-volatile

memory.

713

Bad readback cal data

The specified readback calibration values (

CAL:VOLT

or

CAL:CURR

) are out

of range. Note that the new calibration constants are not stored in the non-

volatile memory.