Chapter 2 Initial Operation

Output Checkout

29

2

Output Checkout

The following procedures check to ensure that the power supply develops its

rated outputs and properly responds to operation from the front panel. For

complete performance and verification tests, refer to the

Service Guide

.

For each step, use the keys shown on the left margins.

Voltage Output Checkout

The following steps verify basic voltage functions with no load.

1 Turn on the power supply.

The power supply will go into the

power-on / reset

state; the output is disabled

(the

OFF

annunciator turns on); the 15V/7A range is selected (the

15V

annunciator turns on); and the knob is selected for

voltage

control.

2 Enable the outputs.

The

OFF

annunciator turns off and the

15V

,

OVP

,

OCP

, and

CV

annunciators are

lit. The

blinking

digit can be adjusted by turning the knob. Notice that the

display is in the meter mode. ‘‘Meter mode’’ means that the display shows the

actual output voltage and current.

3

Check that the front-panel voltmeter properly responds to knob

control for the 15V/7A range.
Turn the knob clockwise or counter clockwise to check that the voltmeter

responds to knob control and the ammeter indicates nearly zero.

4

Ensure that the voltage can be adjusted from zero to the full rated

value.

1

Adjust the knob until the voltmeter indicates 0 volts and then adjust the knob

until the voltmeter indicates ‘‘

15.0

volts’’.

1

You can use the resolution selection keys to move the blinking digit to the

right or left when setting the voltage.