Connection to a Computer or Terminal

To connect the power supply to a computer or terminal, you must have the

proper interface cable. Most computers and terminals are DTE (

Data Terminal

Equipment

) devices. Since the power supply is also a DTE device, you must

use a DTE-to-DTE interface cable. These cables are also called

null-modem,

modem-eliminator

, or

crossover cables

.

The interface cable must also have the proper connector on each end and the

internal wiring must be correct. Connectors typically have 9 pins (DB-9

connector) or 25 pins (DB-25 connector) with a “male” or “female” pin

configuration. A male connector has pins inside the connector shell and a

female connector has holes inside the connector shell.

If you cannot find the correct cable for your configuration, you may have to

use a

wiring adapter

. If you are using a DTE-to-DTE cable, make sure the

adapter is a “straight-through” type. Typical adapters include gender changers,

null-modem adapters, and DB-9 to DB-25 adapters.

The cable and adapter diagrams shown below can be used to connect the

power supply to most computers or terminals

. If your configuration is

different than those described, order the Agilent 34399A

Adapter Kit

. This kit

contains adapters for connection to other computers, terminals, and modems.

Instructions and pin diagrams are included with the adapter kit.

DB-9 Serial Connection

If your computer or terminal has a 9-pin serial port

with a male connector, use the null-modem cable included with the

Agilent

34398A Cable Kit.

This cable has a 9-pin female connector on each end. The

cable pin diagram is shown below

.

Instrument

PC

DB9

Male

DB9

Female

DB9

Female

DB9

Male

DCD

RX
TX

DTR

1

2
3

4

5
6

7

8
9

GND

DSR

RTS
CTS

RI

1

2

3
4

5

6
7

8

9

DCD
RX

TX

DTR

GND

DSR
RTS

CTS

RI