Rs-232 troubleshooting 65, Rs-232 troubleshooting – Agilent Technologies E3632A User Manual

Page 67

Advertising
background image

Chapter 3 Front-Panel Operation

RS-232 Interface Configuration

65

3

2

When the power supply wants to “talk” over the interface (which means that

it has processed a query) and has received a

<new line>

message terminator,

it will set the DTR line FALSE. This implies that once a query has been sent to

the power supply, the bus controller should read the response before

attempting to send more data. It also means that a

<new line>

must terminate

the command string. After the response has been output, the power supply sets

the DTR line TRUE again, unless the first case (see above) prevents this.

The power supply monitors the DSR line to determine when the bus controller

is ready to accept data over the interface. The power supply monitors the DSR

line (pin 6 on the RS-232 connector) before each character is sent. The output

is suspended if the DSR line is FALSE. When the DSR line goes TRUE,

transmission will resume.

The power supply holds the DTR line FALSE while output is suspended. A form

of interface

deadlock

exists until the bus controller asserts the DSR line TRUE

to allow the power supply to complete the transmission. You can break the

interface deadlock by sending the

<Ctrl-C>

character, which clears the

operation in progress and discards pending output (this is equivalent to the

IEEE-488 device clear action).

For the <Ctrl-C> character to be recognized reliably by the power supply

while it holds DTR FALSE, the bus controller must first set DSR FALSE.

RS-232 Troubleshooting

Here are a few things to check if you are having problems communicating over

the RS-232 interface. If you need additional help, refer to the documentation

that came with your computer.

• Verify that the power supply and your computer are configured for the same

baud rate, parity, and number of data bits. Make sure that your computer is

set up for

1 start

bit and

2 stop bits

(these values are fixed on the power

supply).

• Make sure to execute the

SYSTem:REMote

command to place the power

supply in the remote mode.

• Verify that you have connected the correct interface cable and adapters.

Even if the cable has the proper connectors for your system, the internal

wiring may be incorrect. The

Agilent Technologies 34398A Cable Kit

can

be used to connect the power supply to most computers or terminals.

• Verify that you have connected the interface cable to the correct serial port

on your computer (COM1, COM2, etc).