Chapter 7 Tutorial

Connecting the Load

147

7

Connecting the Load

Output Isolation

The output of the power supply is isolated from chassis ground. Any output

terminal may be grounded, or an external voltage source may be connected

between any terminal output and ground. However, output terminals must be

kept within ±60 Vdc when metal shorting bars without insulation are used to

connect the (+) output to the (+) sense and the (-) output and the (-) sense

terminals or ±240 Vdc of ground when metal shorting bars without insulation

are either replaced with insulated conductors or they are removed from the

terminals so there is no operator access to the output conductors without

insulation. A chassis ground terminal is provided on the front panel for

convenience.

Multiple Loads

When connecting multiple loads to the power supply, each load should be

connected to the output terminals using separate connecting wires. This

minimizes mutual coupling effects between loads and takes full advantage of

the low output impedance of the power supply. Each pair of wires should be

as short as possible and twisted or shielded to reduce lead inductance and

noise pick-up. If a shield is used, connect one end to the power supply ground

terminal and leave the other end disconnected.

If cabling considerations require the use of distribution terminals that are

located remotely from the power supply, connect output terminals to the

distribution terminals by a pair of twisted or shielded wires. Connect each load

to the distribution terminals separately.

Table 7-1

. Wire Rating

AWG

10

12

14

16

18

20

22

24

26

28

Suggested

maximum

Current(amps)*

40

25

20

13

10

7

5

3.5

2.5

1.7

m

Ω/ft

1.00

1.59

2.53

4.02

6.39

10.2

16.1

25.7

40.8

64.9

m

Ω/m

3.3

5.2

8.3

13.2

21.0

33.5

52.8

84.3

133.9 212.9

*Single conductor in gree air at 30 °C with insulation