Programming overcurrent protection, 1 turn on the power supply, 2 enable the output – Agilent Technologies E3632A User Manual

Page 47: 3 enter the ocp menu and set the trip level, 4 enable the ocp circuit, Programming overcurrent protection 45, Setting the ocp level and enable the ocp circuit

Advertising
background image

Chapter 3 Front-Panel Operation

Programming Overcurrent Protection

45

3

Programming Overcurrent Protection

Overcurrent protection guards the load against output currents that reach a

specified value greater than the programmed protection level. It is

accomplished by programming the output current to zero.

The following steps show how to set the overcurrent protection trip level, how

to check OCP operation and how to clear overcurrent condition.

• Front-panel operation:

Setting the OCP Level and Enable the OCP Circuit

1

Turn on the power supply.

The power supply will go into the

power-on / reset

state; the output is disabled

(the

OFF

annunciator turns on); the 15V/7A range is selected (the

15V

annunciator turns on); and the knob is selected for

voltage

control.

2

Enable the output.
The

OFF

annunciator turns off and the display will go to the meter mode.

3

Enter the OCP menu and set the trip level.

You will see the above message on the display when you enter the OCP menu.

Adjust the knob for the desired OCP trip level.

4

Enable the OCP circuit.

You will see the above message after pressing the

key.

LEVEL 7.5 A

OCP ON

Power

Output On/Off

Over Current

Over Current

Over Current

Advertising