Chapter 7 Tutorial

Connecting the Load

150

Reverse Current Loading

An active load connected to the supply may actually deliver a reverse current

to the supply during a portion of its operating cycle. An external source can

not be allowed to pump current into the supply without risking loss of

regulation and possible damage. These effects can be avoided by preloading

the output with a dummy load resistor. The dummy load resistor should draw

at least the same amount of current from the supply as the active load may

deliver to the supply. The value of the current for the dummy load plus the

value of the current the load draws from the supply must be less than the

maximum current of the supply.