Chapter 3 Front-Panel Operation

System-Related Operations

52

System-Related Operations

This section gives information on topics such as self-test, error conditions, and

front-panel display control. This information is not directly related to setting

up the power supply but is an important part of operating the power supply.

Self-Test

A

power-on

self-test occurs automatically when you turn on the power supply.

This test assures you that the power supply is operational. This test does not

perform the extensive set of tests that are included as part of the complete self-

test described below. If the power-on self-test fails, the

ERROR

annunciator

turns on.

• A

complete

self-test performs a series of tests and takes approximately 2

seconds to execute. If all tests pass, you can have a high confidence that the

power supply is operational.

• If the

complete

self-test is successful, “PASS” is displayed on the front panel.

If the self-test fails, “FAIL” is displayed and the

ERROR

annunciator turns on.

See the

Service Guide

for instructions on returning the power supply to Agilent

Technologies for service.

• Front-panel operation:

The

complete

self-test is enabled by pressing the

the(actually any front

panel keys except the

key) and the power-line switch simultaneously

and then continuing to press the

key for 5 seconds. The complete self-

test will be finished in 2 seconds.

• Remote interface operation:

*TST?

Returns “0” if the complete self-test passes or “1” if it fails.

Recall

Error

Recall