Chapter 5 Error Messages

Calibration Errors

130

714

Bad OVP cal data

The overvoltage protection calibration constant is out of range. Note that the

new calibration constants are not stored in the non-volatile memory.

715

Bad OCP cal data

The overcurrent protection calibration constant is out of range. Note that the

new calibration constants are not stored in the non-volatile memory.

716

Bad DAC DNL error correction data

Invalid data measured during the calibration for DAC differential nonlinearity

error correction.

717

Cal OVP or OCP status enabled

Overvoltage protection status or overcurrent protection status is enabled. You

must set both overvoltage and overcurrent protection status to OFF before and

during the calibration.

740

Cal checksum failed, secure state

741

Cal checksum failed, string data

742

Cal checksum failed, store/recall data in location 0

743

Cal checksum failed, store/recall data in location 1

744

Cal checksum failed, store/recall data in location 2

745

Cal checksum failed, store/recall data in location 3

746

Cal checksum failed, DAC cal constants

747

Cal checksum failed, readback cal constants

748

Cal checksum failed, GPIB address

749

Cal checksum failed, internal data

750

Cal checksum failed, DAC DNL error correction data