Chapter 3 Front-Panel Operation

Disabling the Output

50

Disabling the Output

The output of the power supply can be disabled or enabled from the front panel.

• When the power supply is in the “Off” state, the

OFF

annunciator turns on and

the output is disabled. The

OFF

annunciator turns off when the power supply

returns to the “On” state. When the output is disabled, the voltage value is 0

volts and the current value is 0.02 amps.

• The output state is stored in

volatile

memory; the output is always disabled

when power has been off or after a remote interface reset.

• Front-panel operation:

You can disable the output by pressing

key. This key toggles

between output “Off” and “On” states.

• Remote interface operation:

OUTP {OFF|ON}

Disable or enable the output

While the output is disabled, the range selection keys, the control knob,

resolution selection keys, and adjust selection key are still working. If the

display is in the meter mode, you cannot see the changes of output

voltage and current settings on the display when turning the knob. To see

or check the changes when the outputs are disabled, the display should be

in the limit mode.

Output On/Off