Agilent Technologies E3632A User Manual

Page 129

Advertising
background image

Chapter 5 Error Messages

Execution Errors

126

-221

Settings conflict

Indicates that a legal program data element was parsed but could not be

executed due to the current device state.

-222

Data out of range

A numeric parameter value is outside the valid range for the command.

Example:

TRIG:DEL -3

-223

Too much data

A character string was received but could not be executed because the string

length was more than 40 characters. This error can be generated by the

CALibration:STRing

command.

-224

Illegal parameter value

A discrete parameter was received which was not a valid choice for the

command. You may have used an invalid parameter choice.

Example:

DISP:STAT XYZ

(

XYZ

is not a valid choice).

-330

Self-test failed

The power supply’s complete self-test failed from the remote interface (

*TST?

command). In addition to this error, more specific self-test errors are also

reported.

See also ‘‘Self-Test Errors’’, starting on page 128

.

-350

Too many errors

The error queue is full because more than 20 errors have occurred. No

additional errors are stored until you remove errors from the queue. The error

queue is cleared when power has been off, or after a

*CLS

(clear status)

command has been executed.

-410

Query INTERRUPTED

A command was received which sends data to the output buffer, but the output

buffer contained data from a previous command (the previous data is not

overwritten). The output buffer is cleared when power has been off, or after a

*RST

(reset) command has been executed.

-420

Query UNTERMINATED

The power supply was addressed to talk (i.e., to send data over the interface)

but a command has not been received which sends data to the output buffer.

For example, you may have executed an

APPLy

command (which does not

generate data) and then attempted an

ENTER

statement to read data from the

remote interface.