Chapter 5 Error Messages

Execution Errors

127

5

-430

Query DEADLOCKED

A command was received which generates too much data to fit in the output

buffer and the input buffer is also full. Command execution continues but all

data is lost.

-440

Query UNTERMINATED after indefinite response

The

*IDN?

command must be the last query command within a command

string.

Example:

*IDN?;:SYST:VERS?

501

Isolator UART framing error

502

Isolator UART overrun error

511

RS-232 framing error

512

RS-232 overrun error

513

RS-232 parity error

514

Command allowed only with RS-232

There are three commands which are only allowed with the RS-232 interface:

SYSTem:LOCal

,

SYSTem

:

REMote

, and

SYSTem:RWLock

.

521

Input buffer overflow

522

Output buffer overflow

550

Command not allowed in local

You should always execute the

SYSTem:REMote

command before sending

other commands over the RS-232 interface.