Chapter 3 Front-Panel Operation

Constant Current Operation

39

3

5

Adjust the knob for the desired output current.

1

Check that the

Limit

annunciator still blinks. Set the knob for

current

control.

The second digit of the ammeter will be

blinking

. Change the blinking digit

using the resolution selection keys and adjust the knob to the desired output

current.

6

Return to the meter mode.
Press

key or let the display time-out after several seconds to

return the meter mode. Notice that the

Limit

annunciator turns off and the

display shows “OUTPUT OFF” message.

7

Enable the output.
The

OFF

annunciator turns off and the

15V

or

30V

,

OVP

,

OCP

and

CC

annunciators are lit. Notice that the display is in the meter mode. In the

meter

mode, the display shows the actual output voltage and current.

Refer to “Programming Overvoltage Protection” and “Programming

Overcurrent Protection” sections, starting on page 42 and page 45 for more

information on

OVP

and

OCP

annunciators.

8

Verify that the power supply is in the constant current mode.
If you operate the power supply in the constant current (CC) mode, verify that

the

CC

annunciator is lit. If the

CV

annunciator is lit, choose a

higher

voltage

limit.

N o t e

During actual CC operation, if a load change causes the voltage limit to be

exceeded, the power supply will automatically crossover to constant voltage

mode at the preset voltage limit and the output current will drop

proportionately.

• Remote interface operation:

VOLTage {<voltage>|MIN|MAX}

Set the voltage

CURRent {<current>|MIN|MAX}

Set the current

OUTPut ON

Enable the output

1

You can use the resolution selection keys to move the blinking digit to the

right or left when setting the current.