Chapter 7 Tutorial

Connecting the Load

148

W a r n i n g

To satisfy safety requirements, load wires must be heavy enough not to

overheat while carrying the short-circuit output current of the power supply.

Remote Voltage Sensing

Normally, a power supply operating in the constant voltage mode achieves its

optimum line and load regulations, its lowest output impedance, drift, and

ripple and noise, and its fastest transient recovery performance at the power

supply output terminals. If the load is separated from the output terminals by

any lead length, some of these performance characteristics will be degraded

at the load terminals - usually by an amount proportional to the impedance of

the load leads compared with the output impedance of the power supply.

With remote voltage sensing, a feature included in the Agilent E3632A power

supply, it is possible to connect the input of the voltage feedback amplifier

directly to the load terminals so that the regulator performs its function with

respect to the load terminals rather than with respect to the power supply

output terminals. Thus, the voltage at the power supply output terminals shifts

by whatever amount is necessary to compensate for the voltage drop in the

load leads, thereby maintaining the voltage at the load terminals constant.

Figure 7-7. Regulated Power Supply with Remote Sensing