Chapter 3 Front-Panel Operation

Calibration Overview

70

Calibration Count

You can determine the number of times that your power supply has been

calibrated. Your power supply was calibrated before it left the factory. When

you receive your power supply, read the count to determine its initial value.

The calibration count feature can be performed from the remote interface

only.

• The calibration count is stored in

non-volatile

memory, and

does not

change

when power has been off or after a remote interface reset.

• The calibration count increments up to a maximum of 32,767 after which it

wraps-around to 0. Since the value increments by one for each calibration

point, a complete calibration will increase the value by 5 counts.

• Remote Interface Operation:

CAL:COUN?

Query the number of times of calibration

Calibration Message

You can use the calibration message feature to record calibration information

about your power supply. For example, you can store such information as the

last calibration date, the next calibration due date, the power supply’s serial

number, or even the name and phone number of the person to contact for a

new calibration.

You can record and read information in the calibration message from the

remote interface only.

• The power supply should be unsecured before sending a calibration

message.

• The calibration message may contain up to 40 characters.

• The calibration message is stored in

non-volatile

memory, and

does not

change when power has been off or after a remote interface reset.

• Remote Interface Operation:

CAL:STR <quoted string> Store the cal message

The following command string shows how to store a calibration message.

"CAL:STR ‘CAL 05-1-97’ "