Chapter 4 Remote Interface Reference

RS-232 Interface Commands

99

4

RS-232 Interface Commands

Use the front-panel ‘‘I/O Config’’ key to select the baud rate, parity, and the

number of data bits (see chapter 3 ‘‘Remote Interface Configuration’’,

starting on page 56).

SYSTem:LOCal

This command places the power supply in the local mode during RS-232

operation. All keys on the front panel are fully functional.

SYSTem:REMote

This command places the power supply in the remote mode for RS-232

operation. All keys on the front panel, except the ‘‘Local’’ key, are disabled.

SYSTem:RWLock

This command places the power supply in the remote mode for RS-232

operation. This command is the same as the

SYST:REM

command except that

all keys on the front panel are disabled, including the ‘‘Local’’ key.

Ctrl-C

This command clears the operation in progress over the RS-232 interface and

discards any pending output data.

This is equivalent to the IEEE-488 device

clear action over the GPIB interface.

It is very important that you send the

SYST:REM

command to place the

power supply in the remote mode. Sending or receiving data over the RS-

232 interface when not configured for remote operation can cause

unpredictable results.