Chapter 4 Remote Interface Reference

Using the APPLy Command

81

4

Using the APPLy Command

The

APPLy

command provides the most straightforward method to program

the power supply over the remote interface. You can select the output voltage

and current in one command.

APPLy {<voltage>| DEF | MIN | MAX}[,{<current>| DEF | MIN | MAX}]

This command is combination of

VOLTage

and

CURRent

commands. As long

as the newly programmed values are within the presently selected range, the

output voltage and current are changed as soon as the command is executed.

The

APPLy

command changes the power supply’s output to the newly

programmed values only if the programmed values are valid within the

presently selected range. An execution error will occur if the programmed

values are not valid within the selected range.

You can substitute ‘‘MINimum’’, ‘‘MAXimum’’, or ‘‘DEFault’’ in place of a

specific value for the

voltage

and

current

parameters. MIN selects the lowest

values of ‘‘0’’ volts and ‘‘0’’ amps. MAX selects the highest values allowed for

the selected range.

The

default

values of voltage and current are ‘‘0’’ volts and ‘‘7’’ amps regardless

of the presently selected range.

See Table 4-1 for details of parameters.

If you specify only one parameter of the

APPLy

command, the power supply

regards it as voltage setting value.

APPLy?

This command queries the power supply’s present

voltage

and

current

setting

values and returns a quoted string. The voltage and current are returned in

sequence as shown in the sample string below (the quotation marks are

returned as part of the string).

"15.00000,4.00000"

In the above string, the first number 15.00000 is the voltage setting value and

the second number 4.00000 is the current setting value.