Chapter 3 Front-Panel Operation

Constant Voltage Operation

37

3

5

Adjust the knob for the desired output voltage.

1

Check that the

Limit

annunciator still blinks. Set the knob for

voltage

control.

The second digit of the voltmeter will be

blinking

. Change the blinking digit

using the resolution selection keys and adjust the knob to the desired output

voltage.

6

Return to the meter mode.
Press

key or let the display time-out after several seconds to

return to the meter mode. Notice that the

Limit

annunciator turns off and the

display shows “OUTPUT OFF” message.

7

Enable the output.
The

OFF

annunciator turns off and the

15V

or

30V

,

OVP

,

OCP

and

CV

annunciators are lit. Notice that the display is in the meter mode. In the

meter

mode, the display shows the actual output voltage and current.

Refer to “Programming Overvoltage Protection” and “Programming

Overcurrent Protection” sections, starting on page 42 and page 45 for more

information on

OVP

and

OCP

annunciators.

8

Verify that the power supply is in the constant voltage mode.
If you operate the power supply in the constant voltage (CV) mode, verify that

the

CV

annunciator is lit. If the

CC

annunciator is lit, choose a

higher

current

limit.

N o t e

During actual CV operation, if a load change causes the current limit to be

exceeded, the power supply will automatically crossover to the constant

current mode at the preset current limit and the output voltage will drop

proportionately.

• Remote interface operation:

CURRent {<current>|MIN|MAX}

Set the current

VOLTage {<voltage>|MIN|MAX

Set the voltage

OUTPut ON

Enable the output

1

You can use the resolution selection keys to move the blinking digit to the

right or left when setting voltage.

Volt/Curr

Display Limit

Display Limit

Output On/Off