Output Characteristics

146

Figure 7-6. Simplified Diagram of Common Mode and Normal Mode Sources of Noise

When the load changes very rapidly, as when a relay contact is closed, the

inductance in the hook up wire and in the power supply output will cause a

spike to appear at the load. The spike is a function of the rate of change of the

load current. When very rapid changes in load are expected, a capacitor with

a low series resistance, in parallel with the power supply, and close to the load

is the best way to minimize these voltage spikes.