Chapter 8 Specifications

Supplemental Characteristics

160

Output Terminal Isolation (maximum, from chassis ground)
±60 Vdc when connecting shorting conductors without insulation to the (+) output to

the (+) sense and the (-) output and the (-) sense terminals.
±240 Vdc when connecting insulated shorting conductors to the (+) output to the (+)

sense and the (-) output and the (-) sense terminals.

AC Input Ratings (selectable via rear panel selector)
std

115 Vac ± 10%, 47 to 63 Hz

opt 0E3

230 Vac ± 10%, 47 to 63 Hz

opt 0E9

100 Vac ± 10%, 47 to 63 Hz

Maximum Input Power
500 VA with full load

Cooling
Fan cooled

Operating Temperature
0 to 40 °C for full rated output. At higher temperatures, the output current is derated

linearly to 50% at 55 °C maximum temperature.

Output Voltage Overshoot
During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V if the

output control is set to less than 1 V. If the output control is set to 1 V or higher, there

is no overshoot.

Programming Language
SCPI (Standard Commands for Programmable Instruments)

State Storage Memory
Three (3) user-configurable stored states

Recommended Calibration Interval
1 year