Programming Overvoltage Protection

42

Programming Overvoltage Protection

Overvoltage protection guards the load against output voltages that reach a

specified value greater than the programmed protection level. It is

accomplished by shorting the output via an internal SCR when the trip level is

set to equal or greater than 3 volts, or by progamming the output to 1 volt when

the trip level is set to less than 3 volts.

The following steps show how to set the OVP trip level, how to check OVP

operation, and how to clear overvoltage condition.

• Front-panel operation:

Setting the OVP Level and Enable the OVP Circuit

1

Turn on the power supply.

The power supply will go into the

power-on / reset

state; the output is disabled

(the

OFF

annunciator turns on); the 15V/7A range is selected (the

15V

annunciator turns on); and the knob is selected for

voltage

control.

2

Enable the output.
The

OFF

annunciator turns off and the display will go to the meter mode.

3

Enter the OVP menu and set the trip level.

You will see the above message on the display when you enter the OVP menu.

Adjust the control knob for the desired OVP trip level.

N o t e

Note that you cannot set the trip levels to lower than 1.0 volt.

4

Enable the OVP circuit.

You will see the above message after pressing

key.

LEVEL 32.0 V

OVP ON

Power

Output On/Off

Over Voltage

Over Voltage

Over Voltage