Output setting and operation commands, Current{<current>|minimum | maximum|up|down, Current – Agilent Technologies E3632A User Manual

Page 85: Output setting and operation commands 82

Advertising
background image

Chapter 4 Remote Interface Reference

Output Setting and Operation Commands

82

Output Setting and Operation Commands

This section describes low-level commands used to program the power supply.

Although the

APPLy

command provides the most straightforward method to

program the power supply, the low-level output setting commands give you

more flexibility to change the individual parameters.

CURRent{<current>|MINimum | MAXimum|UP|DOWN}

This command programs the immediate current level of the power supply. The

immediate level is the current value of the output terminals.

The

CURRent

command changes the output of the power supply to the newly

programmed value regardless of the output range presently selected.

You can substitute ‘‘MINimum’’ or ‘‘MAXimum’’ in place of a specific value for

the current parameter. MIN selects the lowest current values of ‘‘0’’ amps. MAX

selects the highest current values allowed for the selected range.

This command also increases or decreases the immediate current level using

the ‘‘UP’’ or ‘‘DOWN’’ parameter by a predetermined amount. The command

CURRent

:

STEP

sets the amount of increase or decrease. Notice that a new

increment setting will cause an execution error -222 (Data out of range) when

the maximum or the minimum rated current is exceeded.

CURRent

The following program segments show how to use the

CURR

UP

or

CURR

DOWN

Example

command to increase or decrease the output current with the

CURR

:

STEP

command.

"CURR:STEP 0.01"

Set the step size to 0.01 A

"CURR UP"

Increase the output current

"CURR:STEP 0.02"

Set the step size to 0.02 A

"CURR DOWN"

Decrease the output current