5 exit the ocp menu, 2 move to the clear mode, Checking ocp operation 46 – Agilent Technologies E3632A User Manual

Page 48: Clearing the overcurrent condition 46, Checking ocp operation, Clearing the overcurrent condition

Advertising
background image

Chapter 3 Front-Panel Operation

Programming Overcurrent Protection

46

5

Exit the OCP menu.

The “CHANGED” message is displayed for a second to show that the new OCP

trip level is now in effect. If the OCP settings are not changed, “NO CHANGE”

will be displayed. The power supply will exit the OCP menu and the display

will return to the meter mode. Check that the

OCP

annunciator turns on.

Checking OCP Operation

To check OCP operation, raise the output current to near the trip point. Then

very gradually increase the output by turning the knob until the OCP circuit

trips. This will cause the power supply’s output current to drop to zero and the

OCP

annunciator to

blink

. The “OCP TRIPPED” message also appears on the

display.

Clearing the Overcurrent Condition

When the OCP condition occurs (the “OCP TRIPPED” message is shown on

the display), the

OCP

annunciator blinks. When it was caused by external

voltage source such as a battery, disconnect it first. The following steps show

how to clear the overcurrent conditions and get back to normal mode

operation. In the following steps, the display will go back to "OCP TRIPPED"

if you let the display time out after about several seconds.

or

1

Readjust the OCP trip level or the output current level.

Lower the output current level below the OCP trip point after pressing the

or raise the OCP trip level by using the knob after pressing

the

key.

2

Move to the clear mode.

You will see the above message after pressing the

key. If you

changed the output current level, press the

key twice. Turn the

knob to the right until the "OCP CLEAR" appears on the display.

CHANGED

ocp on

Over Current

Over Current
Display Limit

Display Limit

Over Current

Over Current

Over Current