Chapter 8 Specifications

Supplemental Characteristics

159

8

Supplemental Characteristics

Output Programming Range (maximum programmable values)
Low range

0 to 15.45 V/0 to 7.21A

High range

0 to 30.9 V/0 to 4.12 A

OVP

1 V to 32 V

OCP

0 A to 7.5 A

Remote Sensing Capability
Voltage drop

Up to 1 V per each lead

Load regulation Add 5 mV to spec for each 1-volt change in the + output

lead due to load current changes.

Load voltage

Subtract voltage drop in load leads from specified output

voltage rating.

Temperature Coefficient, ±(% of output + offset)
Maximum change in output/readback per °C after a 30-minute warm-up
Voltage

0.01% + 3 mV

Current

0.02% + 3 mA

Stability, ±(% of output + offset)
Following 1 hour warm-up, change in output over 8 hours under constant load, line,

and ambient temperature

Voltage

0.02% + 1 mV

Current

0.1% + 1 mA

Voltage Programming Speed
Maximum time required for output voltage to settle within 1% of its total excursion (for

resistive load). Excludes command processing time.

Full load

No laod

Up

50 msec

20 msec

Down

45 msec

400 msec