Chapter 4 Remote Interface Reference

The SCPI Status Registers

105

4

The Status Byte Summary register is cleared when:

• You execute the

*CLS

(clear status) command.

• Querying the Standard Event register (

*ESR?

command) will clear only bit

5 in the Status Byte summary register.

For example, 24 (8 + 16) is returned when you have queried the status of the

Status Byte register, QUES and MAV conditions have occurred.
The Status Byte Enable register (Request Service) is cleared when:

• You execute the

*SRE 0

command.

• You turn on the power and have previously configured the power supply

using the

*PSC 1

command.

• The enable register

will not

be cleared at power-on if you have previously

configured the power supply using

*PSC 0

.

Using Service Request (SRQ) and Serial POLL

You must configure your bus controller to respond to the IEEE-488 service

request (SRQ) interrupt to use this capability. Use the Status Byte enable

register (

*SRE

command) to select which summary bits will set the low-level

IEEE-488 service request signal. When bit 6 (request service) is set in the Status

Byte, an IEEE-488 service request interrupt message is automatically sent to

the bus controller. The bus controller may then poll the instruments on the bus

to identify which one requested service (the instrument with bit 6 set in its

Status Byte).

The request service bit is cleared only by reading the Status Byte using an

IEEE-488 serial poll or by reading the event register whose summary bit is

causing the service request.

To read the Status Byte summary register, send the IEEE-488 serial poll

message. Querying the summary register will return a decimal value which

corresponds to the binary-weighted sum of the bits set in the register. Serial

poll will automatically clear the “request service” bit in the Status Byte

summary register. No other bits are affected. Performing a serial poll will not

affect instrument throughput.