Performance specifications, Performance specifications 157 – Agilent Technologies E3632A User Manual

Page 160

Advertising
background image

Chapter 8 Specifications

Performance Specifications

157

8

Performance Specifications

Output Ratings(@0°C - 40°C)
Low range

0 to +15 V/0 to 7 A

High range

0 to +30 V/0 to 4 A

Programming Accuracy

[1]

12 months(@25°C ± 5°C), ±(% of output + offset)

Voltage

0.05% + 10 mV

Current

0.2% + 10 mA

Readback Accuracy

[1]

12 months(over GPIB and RS-232 or front panel with respect

to actual output @25°C ± 5°C), ±(% of output + offset)
Voltage

0.05% + 5 mV

Current

0.15% + 5 mA

Ripple and Noise (with outputs ungrounded, or with either output terminal grounded,

20 Hz to 20 MHz)
Normal mode voltage

<0.35 mV rms and 2 mV p-p

Normal mode current

<2 mA rms

Common mode current

<1.5

μA rms

Load Regulation, ±(% of output + offset)
Change in output voltage or current for any load change within ratings with remote

sensing connected
Voltage

<

0.01% + 2 mV

Current

<

0.01% + 250

μA

Line Regulation, ±(% of output + offset)
Change in output voltage and current for any line change within ratings
Voltage

<

0.01% + 2 mV

Current

<

0.01% + 250

μA

[1]

Accuracy specifications are after an 1-hour warm-up with no load and calibration at

25 °C.

Advertising