The standard event register 103, The standard event register, Command – Agilent Technologies E3632A User Manual

Page 106: Event status enable) command

Advertising
background image

Chapter 4 Remote Interface Reference

The SCPI Status Registers

103

4

The Standard Event Register

The Standard Event register reports the following types of instrument events:

power-on detected, command syntax errors, command execution errors, self-

test or calibration errors, query errors, or when an

*OPC

command is executed.

Any or all of these conditions can be reported in the standard event summary

bit (ESB, bit 5) of Status Byte register through the enable register. To set the

enable register mask, you write a decimal value to the register using the

*ESE

(Event Status Enable) command.

Table 4-3

. Bit Definitions – Standard Event Register

An error condition (Standard Event register bit 2, 3, 4, or 5) will always

record one or more errors in the power supply’s error queue. Read the

error queue using the

SYST:ERR?

command.

Bit

Decimal

Value

Definition

0

OPC

1

Operation Complete. All commands prior to and

including an *OPC command have been executed.

1

Not Used

0

Always set to 0.

2

QYE

4

Query Error. The power supply tried to read the output

buffer but it was empty. Or, new command line was

received before a previous query had been read. Or,

both the input and output buffers are full.

3

DDE

8

Device Error. A self-test or calibration error occurred

(see error numbers 601 through 750 in chapter 5).

4

EXE

16

Execution Error. An execution error occurred (see error

numbers -211 through -224 in chapter 5).

5

CME

32

Command Error. A command syntax error occurred

(see error numbers -101 through -178 in chapter 5).

6

Not Used

0

Always set to 0.

7

PON

128

Power On. Power has been turned off and on since the

last time the event register was read or cleared.