Input Power Requirements

22

Input Power Requirements

You can operate your power supply from a nominal 100 V, 115 V, or 230 V single

phase ac power source at 47 to 63 Hz. An indication on the rear panel shows

the nominal input voltage set for the power supply at the factory. If necessary,

you can change the power-line voltage setting according to the instructions on

the next page.

Power-Line Cord

The power supply is shipped from the factory with a power-line cord that has

a plug appropriate for your location. Contact the nearest Agilent Technologies

Sales and Service Office if the wrong power-line cord is included with your

power supply. Your power supply is equipped with a 3-wire grounding type

power cord; the third conductor being the ground. The power supply is

grounded only when the power-line cord is plugged into an appropriate

receptacle. Do not operate your power supply without adequate cabinet

ground connection.

Power-Line Voltage Selection

Power-line voltage selection is accomplished by adjusting two components:

power-line voltage selector and power-line fuse on the power-line module of

the rear panel. To change the power-line voltage, proceed as follows: