Chapter 3 Front-Panel Operation

Calibration Overview

67

3

To Unsecure for Calibration

You can unsecure the power supply for

calibration either from the front panel or over the remote interface.

The power

supply is secured when shipped from the factory, and the security code is set

to “HP003632”.

• Front-Panel Operation:

If the power supply is secured, you will see the above message for one second

by holding the

key for 5 seconds when you turn on the power

supply. To unsecure the power supply, press the

key after the “CAL

MODE” message is displayed in the calibration mode, enter the security code

using the knob and resolution selection keys, and then press the

key.

When you press the

key to save the change, you will see the message

below for one second if the security code is correct. The unsecured setting is

stored in

non-volatile

memory, and

does not

change when power has been off

or after a remote interface reset.

To exit the calibration mode, turn the power

off and on.

Notice that if the security is incorrect, the power supply displays an

"INVALID" message for a second and returns to the code entering mode for

you to enter the correct code.

• Remote Interface Operation:

CAL:SEC:STAT {OFF|ON},<code> Secure or unsecure the power supply

To unsecure the power supply, send the above command with the same code

used to secure. For example,

"CAL:SEC:STAT OFF, HP003632"

SECURED

000000 CODE

UNSECURED

Calibrate

Secure

Secure

Secure