Voltage:protection? {minimum|maximum, Voltage:protection:state {0|1|off|on, Voltage:protection:state – Agilent Technologies E3632A User Manual

Page 90: Voltage:protection:tripped, Voltage:protection:clear, Voltage:range {p15v|p30v||low|high, Voltage:range

Advertising
background image

Chapter 4 Remote Interface Reference

Output Setting and Operation Commands

87

4

VOLTage:PROTection? {MINimum|MAXimum}

This query returns the overvoltage protection trip level presently programmed.

VOLT:PROT? MAX

and

VOLT:PROT? MIN

return the maximum and minimum

programmable overvoltage trip levels.

VOLTage:PROTection:STATe {0|1|OFF|ON}

This command enables or disables the overvoltage protection function. An

overvoltage condition can be cleared with the

VOLT:PROT:CLE

command

after the condition that caused the OVP trip is removed. At

*RST

, this value is

set to ‘‘ON’’.

VOLTage:PROTection:STATe?

This query returns the state of the overvoltage protection function. The

returned parameter is ‘‘0’’ (OFF) or ‘‘1’’ (ON).

VOLTage:PROTection:TRIPped?

This query returns a ‘‘1’’ if the overvoltage protection circuit is tripped and not

cleared or a ‘‘0’’ if not tripped.

VOLTage:PROTection:CLEar

This command causes the overvoltage protection circuit to be cleared. After

this command, the output voltage is restored to the state it was in before the

protection feature occurred and the OVP trip level remains unchanged to the

value presently programmed. Before sending this command, lower the output

voltage below the trip OVP point, or raise the OVP trip level above the output

setting.

VOLTage:RANGe {P15V|P30V||LOW|HIGH}

This command selects an output range to be programmed by the identifier.

When 15V/7A range is selected, the

maximum programmable

voltage and

current are limited to 15.45 volts and 7.21 amps. When 30V/4A range is selected,

the

maximum programmable

voltage and current are limited to 30.09 volts

and 4.12 amps. ‘‘P30V’’ or ‘‘HIGH’’ is the identifier for the 30V/4A range and

‘‘P15V’’ or ‘‘LOW’’ is for the 15V/7A range. At

*RST

, the 15V/7A range is selected.

VOLTage:RANGe?

This query returns the currently selected range. The returned parameter is

‘‘P30V’’ (HIGH) or ‘‘P15V’’ (LOW).

Advertising