Chapter 4 Remote Interface Reference

The SCPI Status Registers

106

C a u t i o n

The IEEE-488 standard does not ensure synchronization between your bus

controller program and the instrument. Use the

*OPC?

command to

guarantee that commands previously sent to the instrument have completed.

Executing a serial poll before a

*RST, *CLS,

or other commands have

completed can cause previous conditions to be reported.

Using *STB? to Read the Status Byte

The

*STB?

(Status Byte query) command is similar to a serial poll but it is

processed like any other instrument command. The

*STB?

command returns

the same result as a serial poll but the “request service” bit (bit 6)

is not

cleared.

The

*STB?

command is not handled automatically by the IEEE-488 bus

interface hardware and will be executed

only

after previous commands have

completed. Polling is not possible using the

*STB?

command. Executing the

*STB?

command does not clear the Status Byte summary register.

Using the Message Available Bit (MAV)

You can use the Status Byte “message available” bit (bit 4) to determine when

data is available to read into your bus controller. The power supply

subsequently clears bit 4

only

after all messages have been read from the

output buffer.

To Interrupt Your Bus Controller Using SRQ

1

Send a device clear message to clear the power supply’s output buffer (e.g.,

CLEAR 705

).

2

Clear the event registers with the

*CLS

(clear status) command.

3

Set up the enable register masks. Execute the

*ESE

command to set up the

Standard Event register and the

*SRE

command for the Status Byte.

4

Send the

*OPC?

(operation complete query) command and enter the result to

ensure synchronization.

5

Enable your bus controller’s IEEE-488 SRQ interrupt.

Advertising